pH/ORP
HPW7000 pH Measurement System
HPW7000 is an electrode and flow-chamber assembly for pH monitoring in high-purity water.
HPW7000 is an electrode and flow-chamber assembly for pH monitoring in high-purity water. Its compact stainless-steel chamber controls sample flow around the measuring and reference electrodes, with temperature sensing included. A flowing reference and shielded cabling support stable measurements in low-conductivity samples, making the system suitable for crowded power-station water-chemistry sampling panels.
Models: HPW7000
Features & Benefits
- Construction
- Stainless-steel flow chamber and flowing reference
- Supports stable high-purity-water pH readings
Applications
- Power-generation water chemistry
Detailed capabilities
What Is It?
HPW7000 measures pH in high-purity water using a compact flowing-reference assembly suitable for sample panels. It interfaces with UDA2182 or DirectLine. The page states a low-conductivity threshold using '<10 mmho/cm'; that unusual source unit is retained as a documentation qualification rather than silently converting it to a different conductivity unit.
How Does It Work?
A stainless flow chamber shapes sample flow around the measuring/reference electrodes, temperature sensor and solution ground. Keyed cover geometry fixes electrode placement after service, while a clamp permits removal and reinstallation. Shielded leads up to 36 feet avoid a mounting-point preamplifier in the described arrangement, and the flowing reference limits offset during calibration.
What Problems Does It Solve?
Low-conductivity water is prone to noise, streaming potentials and drift that complicate conventional pH measurement. Controlled sample flow, shielding and repeatable electrode placement address those effects, supporting stable turbine-water chemistry monitoring. The compact mounting saves panel space and simplifies calibration without claiming immunity to all sampling or wiring errors.
Environmental Conditions
- Downstream Back Pressure
- -10 in W.C. (min), +1 in W.C. (max)
- Conductivity Range
- >0.055 umhos
- Temperature Range
- Assembly: 10 to 45 deg C; Electrodes: 10 to 80 deg C
- Measurement Range
- pH: 4 to 10
- Resistivity
- 18 megohms-cm (demonstrates system stability at high purity)
- Stability
- ± 0.1 pH per week
Technical Specifications
- Process Connections
- 1/4 in Male, Side mounted on flow assembly
Processing & Performance Metrics
- Accuracy/Noise
- ± 0.1 pH/< 0.1 pH
Applications & Compatibility
- Applications
- Measurement and control of water pH to minimize corrosion and scaling in electric utilities, cogeneration, industrial boilers, and steam generators
| Parameter | Published value / configuration |
|---|---|
| Downstream Back Pressure | -10 in W.C. (min), +1 in W.C. (max) |
| Conductivity Range | >0.055 umhos |
| Temperature Range | Assembly: 10 to 45 deg C; Electrodes: 10 to 80 deg C |
| Measurement Range | pH: 4 to 10 |
| Resistivity | 18 megohms-cm (demonstrates system stability at high purity) |
| Stability | ± 0.1 pH per week |
| Process Connections | 1/4 in Male, Side mounted on flow assembly |
| Accuracy/Noise | ± 0.1 pH/< 0.1 pH |
| Applications | Measurement and control of water pH to minimize corrosion and scaling in electric utilities, cogeneration, industrial boilers, and steam generators |
Values retain manufacturer model qualifications. Review Honeywell specifications ↗
Resources
Resources linked from the manufacturer’s page. Confirm the document revision and applicable model before installation or ordering.
Honeywell resource listing ↗Part Number
Published model / variant identifiers. A full ordering code requires the manufacturer configuration.
Published part numbers
4 parts
| Part Number | Description | Enquiry |
|---|---|---|
| 50024077-501 | SANITARY CLAMP, 2.5" HPW7000 | Request quotation |
| 50044238-501 | ORP ELECTRODE 20' CABLE | Request quotation |
| 50044238-502 | ORP ELECTRODE 36' CABLE | Request quotation |
| HPW7001 | HPW7001-HI-PURITY PH MEASUREMENT SYSTEM | Request quotation |