Fire & Life Safety · Building Systems · Industrial Automation [email protected] +880 1347 561 256
Honeywell Process Measurement & Control

pH/ORP

HPW7000 pH Measurement System

HPW7000 is an electrode and flow-chamber assembly for pH monitoring in high-purity water.

HPW7000 is an electrode and flow-chamber assembly for pH monitoring in high-purity water. Its compact stainless-steel chamber controls sample flow around the measuring and reference electrodes, with temperature sensing included. A flowing reference and shielded cabling support stable measurements in low-conductivity samples, making the system suitable for crowded power-station water-chemistry sampling panels.

Models: HPW7000

Features & Benefits

Construction
Stainless-steel flow chamber and flowing reference
  • Supports stable high-purity-water pH readings

Applications

  • Power-generation water chemistry

Detailed capabilities

What Is It?

HPW7000 measures pH in high-purity water using a compact flowing-reference assembly suitable for sample panels. It interfaces with UDA2182 or DirectLine. The page states a low-conductivity threshold using '<10 mmho/cm'; that unusual source unit is retained as a documentation qualification rather than silently converting it to a different conductivity unit.

Manufacturer reference ↗

How Does It Work?

A stainless flow chamber shapes sample flow around the measuring/reference electrodes, temperature sensor and solution ground. Keyed cover geometry fixes electrode placement after service, while a clamp permits removal and reinstallation. Shielded leads up to 36 feet avoid a mounting-point preamplifier in the described arrangement, and the flowing reference limits offset during calibration.

Manufacturer reference ↗

What Problems Does It Solve?

Low-conductivity water is prone to noise, streaming potentials and drift that complicate conventional pH measurement. Controlled sample flow, shielding and repeatable electrode placement address those effects, supporting stable turbine-water chemistry monitoring. The compact mounting saves panel space and simplifies calibration without claiming immunity to all sampling or wiring errors.

Manufacturer reference ↗

Specifications

Environmental Conditions

Downstream Back Pressure
-10 in W.C. (min), +1 in W.C. (max)
Conductivity Range
>0.055 umhos
Temperature Range
Assembly: 10 to 45 deg C; Electrodes: 10 to 80 deg C
Measurement Range
pH: 4 to 10
Resistivity
18 megohms-cm (demonstrates system stability at high purity)
Stability
± 0.1 pH per week

Technical Specifications

Process Connections
1/4 in Male, Side mounted on flow assembly

Processing & Performance Metrics

Accuracy/Noise
± 0.1 pH/< 0.1 pH

Applications & Compatibility

Applications
Measurement and control of water pH to minimize corrosion and scaling in electric utilities, cogeneration, industrial boilers, and steam generators
HPW7000 pH Measurement System — manufacturer page values
ParameterPublished value / configuration
Downstream Back Pressure-10 in W.C. (min), +1 in W.C. (max)
Conductivity Range>0.055 umhos
Temperature RangeAssembly: 10 to 45 deg C; Electrodes: 10 to 80 deg C
Measurement RangepH: 4 to 10
Resistivity18 megohms-cm (demonstrates system stability at high purity)
Stability± 0.1 pH per week
Process Connections1/4 in Male, Side mounted on flow assembly
Accuracy/Noise± 0.1 pH/< 0.1 pH
ApplicationsMeasurement and control of water pH to minimize corrosion and scaling in electric utilities, cogeneration, industrial boilers, and steam generators

Values retain manufacturer model qualifications. Review Honeywell specifications ↗

Resources

Filter documents by type

Resources linked from the manufacturer’s page. Confirm the document revision and applicable model before installation or ordering.

Honeywell resource listing ↗

Part Number

Published model / variant identifiers. A full ordering code requires the manufacturer configuration.

Published part numbers

4 parts

Part NumberDescriptionEnquiry
50024077-501SANITARY CLAMP, 2.5" HPW7000Request quotation
50044238-501ORP ELECTRODE 20' CABLERequest quotation
50044238-502ORP ELECTRODE 36' CABLERequest quotation
HPW7001HPW7001-HI-PURITY PH MEASUREMENT SYSTEMRequest quotation
Manufacturer sources

https://process.honeywell.com/us/en/products/process-instruments/analytical-instruments-and-sensors/ph-orp/hpw7000-ph-measurement-system

Last source check: 2026-11-02

HPW7000 pH Measurement System Image · Manufacturer ownership retained; republication licence not independently confirmed

ABID Automation LTD maintains this independent product reference. Honeywell owns its product names and trademarks. Current manufacturer documentation governs specification, availability and certification.