7774/7774D Insertion/Removal Ball Valve Mounting

7774 and 7774D assemblies hold pH or ORP electrodes behind a ball-valve mounting so they can be withdrawn under process pressure.

7774 and 7774D assemblies hold pH or ORP electrodes behind a ball-valve mounting so they can be withdrawn under process pressure. Options support Meredian or Durafet sensing and different wetted materials. Restraining mechanisms and the support tube manage insertion and removal, making the arrangement suitable where regular sensor access is needed without interrupting the process.

Features & Benefits

Sensor options
Meredian pH/ORP or Durafet pH
  • Enables electrode service without stopping the process

Applications

  • Boiler water
  • Demineralization
  • Reverse osmosis
  • Chemical and pharmaceutical processes

Detailed capabilities

What Is It?

7774/7774D mounts removable pH/ORP electrodes through a ball-valve assembly. The D form uses Durafet, while 7774 uses the stated Meredian combination electrode. Plastic or 316 stainless wetted components address chemical/thermal needs in boiler analysis, water treatment, pharmaceuticals, chemicals, waste and pulp/paper applications, with the exact assembly chosen for the process.

Manufacturer reference ↗

How Does It Work?

An electrode on a support tube seals into the valve assembly before the valve is opened. It then advances into the process and locks in position; removal reverses that sequence. Restraining mechanisms control the movement. The source specifically permits insertion/withdrawal up to 50 psig, which must remain separate from broader language about high-pressure industrial application.

Manufacturer reference ↗

What Problems Does It Solve?

The arrangement permits electrode service without stopping the process within its stated live-removal limit. Accurate pH measurement supports reagent and quality control, while the described Durafet life and gel-refill features reduce replacement costs. The ball valve does not eliminate the need to follow the restraint, sealing and pressure conditions for insertion or removal.

Manufacturer reference ↗

Specifications

Applications & Compatibility

Applications
Boiler Water Analysis, Demineralization, Reverse Osmosis, Pharmaceutical/Chemical Production, Waste Treatment, Pulp and Paper Production

Mechanical & Physical Properties

Wetted Materials
Plastic: CPVC Ball Valve and Extension Tube, Viton, Teflon, EPDM Seals; Stainless Steel: 316 SS Ball Valve and Extension Tube, Viton, Teflon Seals
Compatible Electrodes
Durafet II and Meredian Series Combination (Single-Probe) pH or ORP Electrodes

Environmental Conditions

Temperature Rating
Plastic: Up to 88 deg C (190 deg F) at 50 psig, 100 psig at 50 deg C (120 deg F); Stainless Steel: Up to 130 deg C (266 deg F) depending on electrode
Pressure Ratings
Safe insertion/removal up to 50 psig; Usable to 100 psig
Flow Velocity
Maximum 5 ft/sec (1.5 m/sec)

Wiring / Mechanical & Physical Properties

Removal Allowance
Plastic: ~1423 mm (56 in); Stainless Steel: ~965 mm (38 in)

I/O Capabilities

Purge Port Connection
Plastic Assembly: 1/4 in NPT Female (Plug supplied) for Inline Cleaning/Flushing
Mounting Types
Allows insertion/removal under pressure without process interruption; Insertion/removal ball valve assembly for pH/ORP electrodes
Mounting Connection
Plastic: 1-1/2 in MNPT; Stainless Steel: 1-1/4 in MNPT
Insertion Depth
Plastic: 6 or 12 in Nominal (with Accessory Nipple); Stainless Steel: 6 in Nominal (not Adjustable)

Network & Connectivity Infrastructure

Overall Length
Plastic: 508 mm (20 in); Stainless Steel: 533 mm (21 in)

Technical Specifications

Electrical Connections
FM-approved intrinsically safe options for Div 1 with 7079 Transmitter; With preamplifier: Suitable for Class I, Div 2, Group A-D
7774/7774D Insertion/Removal Ball Valve Mounting — manufacturer page values
ParameterPublished value / configuration
ApplicationsBoiler Water Analysis, Demineralization, Reverse Osmosis, Pharmaceutical/Chemical Production, Waste Treatment, Pulp and Paper Production
Wetted MaterialsPlastic: CPVC Ball Valve and Extension Tube, Viton, Teflon, EPDM Seals; Stainless Steel: 316 SS Ball Valve and Extension Tube, Viton, Teflon Seals
Compatible ElectrodesDurafet II and Meredian Series Combination (Single-Probe) pH or ORP Electrodes
Temperature RatingPlastic: Up to 88 deg C (190 deg F) at 50 psig, 100 psig at 50 deg C (120 deg F); Stainless Steel: Up to 130 deg C (266 deg F) depending on electrode
Pressure RatingsSafe insertion/removal up to 50 psig; Usable to 100 psig
Flow VelocityMaximum 5 ft/sec (1.5 m/sec)
Removal AllowancePlastic: ~1423 mm (56 in); Stainless Steel: ~965 mm (38 in)
Purge Port ConnectionPlastic Assembly: 1/4 in NPT Female (Plug supplied) for Inline Cleaning/Flushing
Mounting TypesAllows insertion/removal under pressure without process interruption; Insertion/removal ball valve assembly for pH/ORP electrodes
Mounting ConnectionPlastic: 1-1/2 in MNPT; Stainless Steel: 1-1/4 in MNPT
Insertion DepthPlastic: 6 or 12 in Nominal (with Accessory Nipple); Stainless Steel: 6 in Nominal (not Adjustable)
Overall LengthPlastic: 508 mm (20 in); Stainless Steel: 533 mm (21 in)
Electrical ConnectionsFM-approved intrinsically safe options for Div 1 with 7079 Transmitter; With preamplifier: Suitable for Class I, Div 2, Group A-D

Values retain manufacturer model qualifications. Review Honeywell specifications ↗

Resources

Filter documents by type

Resources linked from the manufacturer’s page. Confirm the document revision and applicable model before installation or ordering.

Honeywell resource listing ↗

Part Number

Published part numbers

8 parts

Part NumberDescriptionEnquiry
055919FLOW CHAMBER (3/4 IN OPENING)Request quotation
07773MERIDAN ELECTRODE MOUNTINGRequest quotation
07774DDURAFET II ELECTRODE MOUNTINGRequest quotation
07777DDURAFET II ELECTRODE MOUNTINGRequest quotation
07777DVPDURAFET III ELECTRODE W/VARIO PINRequest quotation
07777DVP-00000107777DVP-02-01-0000-000Request quotation
50163353-413CABLE 10 m FOR D-LX 110Request quotation
50163353-511BLINDKLINKNAGEL A4 3.2X8MMRequest quotation
Manufacturer sources

https://process.honeywell.com/us/en/products/process-instruments/analytical-instruments-and-sensors/ph-orp/7774-7774d-insertion-removal-ball-valve-mounting