Fire & Life Safety · Building Systems · Industrial Automation [email protected] +880 1347 561 256
Honeywell Process Measurement & Control

pH/ORP

7777/7777D Immersion/In-Line Glass Mounting

7777-series assemblies combine measuring and reference electrodes with temperature compensation in a Ryton body.

7777-series assemblies combine measuring and reference electrodes with temperature compensation in a Ryton body. Meredian glass or ORP versions support in-line and immersion installation, while the published Durafet II arrangement is for in-line use. Threaded ends and disconnect plugs simplify fitting and maintenance, allowing selection around the sensing technology and installation method.

Features & Benefits

Mounting
Threaded in-line
  • Simplifies electrode installation and access

Applications

  • Industrial pH and ORP monitoring

Detailed capabilities

What Is It?

7777D uses the non-glass Durafet II pH sensor, while 7777 uses Meredian II glass-pH or platinum/gold ORP electrodes. Both combine measuring/reference functions and temperature compensation in a Ryton body. The source restricts Durafet II mountings to inline use, whereas Meredian supports inline and submerged service; the broad page title must not erase this distinction.

Manufacturer reference ↗

How Does It Work?

Three-quarter-inch NPT ends allow pipeline fitting, and Meredian can attach to a pipe for immersion. Compact construction avoids separate special seals/O-rings in the described installation, with disconnect plugs aiding service. Durafet II has a replaceable reference frit and gel; Meredian II uses a permanent gel-filled junction whose annular geometry reduces fouling and process-current effects.

Manufacturer reference ↗

What Problems Does It Solve?

The mounting simplifies installation and electrode replacement while supporting process pH/ORP control. Durafet's low-impedance response remains useful at low temperature and is described without sodium error or ORP interference. Those non-glass sensor advantages should not be assigned to the Meredian glass/ORP variant merely because both use the 7777 mounting family.

Manufacturer reference ↗

Specifications

Technical Specifications

Electrical Connections
Vario Pin, 11 conductor connector, IP68 rated (33 ft immersion for 1000 hrs)

Mechanical & Physical Properties

Compatible Electrodes
Durafet III pH Electrodes (ISFET Solid-State, Non-Glass)
Wetted Materials
Ryton Body, Silicon ISFET Die, Ceramic Reference Junction, Viton Media Seal, Nickel Plated Brass, EPM Reference Frit Seals

Environmental Conditions

Temperature Rating
Cap Adapter Cable: 0 to 70 deg C (32 to 158 deg F); Electrode: -10 to 110 deg C (14 to 230 deg F)
Pressure Ratings
345 kPa (50 psig) at 100 deg C (212 deg F); 690 kPa (100 psig) at 50 deg C (122 deg F)
Operating Range
0 to 14 pH

I/O Capabilities

Mounting Types
Immersion (Submersion) or In-Line (3/4 in MNPT Process Connection)

Network & Connectivity Infrastructure

Cable Options
DirectLine Remote (20’/50’), Cap Adapter (20’/50’), Preamp (20’/50’), UDA2182 (20’/50’)

Applications & Compatibility

Applications
Wastewater Influent/Effluent, Neutralization, Cooling Towers, Process pH Measurements

Dimensions & Weight

Dimensions
220 x 27 mm Diameter (8.65 x 1.05 in)
7777/7777D Immersion/In-Line Glass Mounting — manufacturer page values
ParameterPublished value / configuration
Electrical ConnectionsVario Pin, 11 conductor connector, IP68 rated (33 ft immersion for 1000 hrs)
Compatible ElectrodesDurafet III pH Electrodes (ISFET Solid-State, Non-Glass)
Wetted MaterialsRyton Body, Silicon ISFET Die, Ceramic Reference Junction, Viton Media Seal, Nickel Plated Brass, EPM Reference Frit Seals
Temperature RatingCap Adapter Cable: 0 to 70 deg C (32 to 158 deg F); Electrode: -10 to 110 deg C (14 to 230 deg F)
Pressure Ratings345 kPa (50 psig) at 100 deg C (212 deg F); 690 kPa (100 psig) at 50 deg C (122 deg F)
Operating Range0 to 14 pH
Mounting TypesImmersion (Submersion) or In-Line (3/4 in MNPT Process Connection)
Cable OptionsDirectLine Remote (20’/50’), Cap Adapter (20’/50’), Preamp (20’/50’), UDA2182 (20’/50’)
ApplicationsWastewater Influent/Effluent, Neutralization, Cooling Towers, Process pH Measurements
Dimensions220 x 27 mm Diameter (8.65 x 1.05 in)

Values retain manufacturer model qualifications. Review Honeywell specifications ↗

Resources

Filter documents by type

Resources linked from the manufacturer’s page. Confirm the document revision and applicable model before installation or ordering.

Honeywell resource listing ↗

Part Number

Published part numbers

5 parts

Part NumberDescriptionEnquiry
51205554-501DURAFET II NPT 8550 VITON SMOORequest quotation
CCDContacting Conductivity SensorsRequest quotation
PHG-1GLASS PH SENSORRequest quotation
PHG-GGLASS ORP SENSOR - GOLDRequest quotation
PHG-PGLASS ORP SENSOR - PLATINUMRequest quotation
Manufacturer sources

https://process.honeywell.com/us/en/products/process-instruments/analytical-instruments-and-sensors/ph-orp/7777-7777d-immersion-in-line-glass-mounting

Last source check: 2026-11-02

7777/7777D Immersion/In-Line Glass Mounting Image · Manufacturer ownership retained; republication licence not independently confirmed

ABID Automation LTD maintains this independent product reference. Honeywell owns its product names and trademarks. Current manufacturer documentation governs specification, availability and certification.

Your next project. A clear starting point.