Fire & Life Safety · Building Systems · Industrial Automation [email protected] +880 1347 561 256
Honeywell Process Measurement & Control

Portable Gauging & Sampling for Terminals

HERMetic UTImeter Gtex 2000 - Closed Type - Land Applications

Gtex 2000 performs closed portable gauging in terminal and refinery tanks, combining ullage, temperature and oil-water interface sensing with manual reference-height checks.

Gtex 2000 performs closed portable gauging in terminal and refinery tanks, combining ullage, temperature and oil-water interface sensing with manual reference-height checks. Its weighted probe and chemical-resistant sealing arrangement suit the published storage duties. Connection to a compatible vapor-control valve contains tank gas, while tape protection and accessible battery replacement support repeated inventory and custody-transfer checks.

Features & Benefits

Checks
Ullage, temperature, interface and manual reference height
  • Contains tank vapors during gauging

Applications

  • Onshore custody transfer
  • Inventory control
  • Free-water detection

Detailed capabilities

Product and configuration

Gtex 2000 performs closed portable gauging in terminal and refinery tanks, combining ullage, temperature and oil-water interface sensing with manual reference-height checks. Its weighted probe and chemical-resistant sealing arrangement suit the published storage duties. Connection to a compatible vapor-control valve contains tank gas, while tape protection and accessible battery replacement support repeated inventory and custody-transfer checks.

  • Checks: Ullage, temperature, interface and manual reference height

Manufacturer reference ↗

Operation and integration

Gtex 2000 combines electronic ullage, temperature and oil-water interface sensing with manual detection of the datum plate for innage/reference-height work. The page specifies a chemical-oriented gasket/connector arrangement and a two-inch load on the one-inch sensing probe. RTD characteristics are stored in the sensor so changing a tape or sensor does not require new system calibration.

Manufacturer reference ↗

Application and selection

A suitable vapor-control valve maintains closed shore-tank operation. Interchangeable adaptors, tape protection and battery access support repeated inventory and custody checks. The source lists historical standards and approvals, which should be associated with the actual selected instrument rather than treated as current blanket acceptance of every tank or medium.

Manufacturer reference ↗

Case studies & further resources

Manufacturer Document

Portable Gauging Devices and Samplers for Land Applications

View resource ↗
Manufacturer Document

Honeywell Authorized Service Stations - HERMetic Portable Gauging & Sampling Products

View resource ↗

Specifications

Mechanical & Physical Properties

Probe
2 in load on sensing probe for operation in any kind of product; Heavy probe for manual dip plate detection

Standards & Certifications

Hazardous Location Approvals
ATEX II 1 G EEx ia IIB T4/Tamb 50 deg C; China CQST ExiaIIBT4; Factory Mutual CL I, DIV 1, CandD, T4 Tamb 50 deg C and CL I, ZN 0, AEx ia IIB T4 Tamb 50 deg C; IECEx Zone 0 Ex ia IIB T4/Tamb 50 deg C
Approvals
API MPMS Chap 3.1A, 7, 8.1; EC Directives; IP PMM Part III-1, IV; ISO 4512, 3170, 4268

Pressure/Operating Limits

Maximum Tank Overpressure
0.3 bar (4.4 psi)

Technical Specifications

Temperature Resolution
0.01 or 0.1 deg selectable; deg C or deg F selectable

I/O Capabilities

Measurement Functions
Ullage (distance from liquid surface to tank top), Temperature, Oil-Water Interface Level, Innage/Reference Height (manual detection of dip datum plate)
Indication
Audible and Visible Selectable
Operation
Avoids gas release during operation; Used with HERMetic vapor control valve

Processing & Performance Metrics

Temperature Accuracy
±0.1 deg C (0 to 70 deg C), ±0.2 deg F (32 to 158 deg F)
Measurement Accuracy
Ullage/Interface: ±2 mm (±0.08 in approx.), Tape Accuracy: ±1.5 mm/30 m (±1/16 in/100 ft)

Environmental Conditions

Temperature Range
Sensor: -40 to 90 deg C (-40 to 194 deg F); Ambient: -20 to 50 deg C (-4 to 122 deg F)

Overview

Special Features
Tape cleaning device, window wiper, tape protection, easy battery and sensor exchange, no calibration needed after sensor exchange Millimeter-PRD

Electrical Specifications

Power Supply
Approved 9V Batteries

Applications & Compatibility

Applications
Custody Transfer, Inventory Control, Free Water Detection on Shore Tanks
Sensor Type
ULTRA Sensing Probe: Stainless Steel Tube, PEEK Head, Ultrasonic Liquid Level Sensor, Temperature Sensor (RTD), Conductivity Electrode

Network & Connectivity Infrastructure

Tape Length
1 mm, 1/16 in resolution; 30 m/100 ft (Standard); Metric/English Graduation

Software & Interface

Display
8-Character LCD with Backlight
HERMetic UTImeter Gtex 2000 - Closed Type - Land Applications — manufacturer page values
ParameterPublished value / configuration
Probe2 in load on sensing probe for operation in any kind of product; Heavy probe for manual dip plate detection
Hazardous Location ApprovalsATEX II 1 G EEx ia IIB T4/Tamb 50 deg C; China CQST ExiaIIBT4; Factory Mutual CL I, DIV 1, CandD, T4 Tamb 50 deg C and CL I, ZN 0, AEx ia IIB T4 Tamb 50 deg C; IECEx Zone 0 Ex ia IIB T4/Tamb 50 deg C
ApprovalsAPI MPMS Chap 3.1A, 7, 8.1; EC Directives; IP PMM Part III-1, IV; ISO 4512, 3170, 4268
Maximum Tank Overpressure0.3 bar (4.4 psi)
Temperature Resolution0.01 or 0.1 deg selectable; deg C or deg F selectable
Measurement FunctionsUllage (distance from liquid surface to tank top), Temperature, Oil-Water Interface Level, Innage/Reference Height (manual detection of dip datum plate)
IndicationAudible and Visible Selectable
OperationAvoids gas release during operation; Used with HERMetic vapor control valve
Temperature Accuracy±0.1 deg C (0 to 70 deg C), ±0.2 deg F (32 to 158 deg F)
Measurement AccuracyUllage/Interface: ±2 mm (±0.08 in approx.), Tape Accuracy: ±1.5 mm/30 m (±1/16 in/100 ft)
Temperature RangeSensor: -40 to 90 deg C (-40 to 194 deg F); Ambient: -20 to 50 deg C (-4 to 122 deg F)
Special FeaturesTape cleaning device, window wiper, tape protection, easy battery and sensor exchange, no calibration needed after sensor exchange Millimeter-PRD
Power SupplyApproved 9V Batteries
ApplicationsCustody Transfer, Inventory Control, Free Water Detection on Shore Tanks
Sensor TypeULTRA Sensing Probe: Stainless Steel Tube, PEEK Head, Ultrasonic Liquid Level Sensor, Temperature Sensor (RTD), Conductivity Electrode
Tape Length1 mm, 1/16 in resolution; 30 m/100 ft (Standard); Metric/English Graduation
Display8-Character LCD with Backlight

Values retain manufacturer model qualifications. Review Honeywell specifications ↗

Resources

Filter documents by type

Resources linked from the manufacturer’s page. Confirm the document revision and applicable model before installation or ordering.

Honeywell resource listing ↗

Part Number

Published part numbers

4 parts

Part NumberDescriptionEnquiry
TS 11289Carter Gtex 2000 baseRequest quotation
TS 12113Carter Gtex 2000Request quotation
TS 40004Hex nut M4Request quotation
TS NGU2KHERMETIC UTIMETER GTEX2000 N SERIESRequest quotation
Manufacturer sources

https://process.honeywell.com/us/en/products/terminals/portable-gauging-and-sampling-for-terminals/hermetic-utimeter-gtex-2000-closed-type-land-applications

Last source check: 2026-11-02

HERMetic UTImeter Gtex 2000 - Closed Type - Land Applications · Manufacturer ownership retained; republication licence not independently confirmed

HERMetic UTImeter Gtex 2000 - Closed Type - Land Applications · Manufacturer ownership retained; republication licence not independently confirmed

HERMetic UTImeter Gtex 2000 - Closed Type - Land Applications · Manufacturer ownership retained; republication licence not independently confirmed

HERMetic UTImeter Gtex 2000 - Closed Type - Land Applications · Manufacturer ownership retained; republication licence not independently confirmed

ABID Automation LTD maintains this independent product reference. Honeywell owns its product names and trademarks. Current manufacturer documentation governs specification, availability and certification.

Your next project. A clear starting point.

Tell us about your facility and what you need to achieve.

CallSend enquiry