Fire & Life Safety · Building Systems · Industrial Automation [email protected] +880 1347 561 256
Honeywell Process Measurement & Control

Fixed Gauging

Lidec Liquid Level Switch

Lidec detects liquid using a surface wave in a stainless-steel probe rather than mechanical movement.

Lidec detects liquid using a surface wave in a stainless-steel probe rather than mechanical movement. Self-monitoring signals instrument faults, while changes in fluid conductivity, density or similar properties do not require repeated calibration. The switches can connect to a ship's alarm system or CARLA, supporting high-level, service-tank, bilge and water-ingress detection duties.

Features & Benefits

Principle
Static surface-wave sensing
Diagnostics
Self-monitoring fault detection
  • Avoids moving-part wear and routine fluid-dependent calibration

Applications

  • Tanker high-level alarms
  • Bilge accumulation
  • Water ingress
  • Service tanks

Detailed capabilities

Product and configuration

Lidec detects liquid using a surface wave in a stainless-steel probe rather than mechanical movement. Self-monitoring signals instrument faults, while changes in fluid conductivity, density or similar properties do not require repeated calibration. The switches can connect to a ship's alarm system or CARLA, supporting high-level, service-tank, bilge and water-ingress detection duties.

  • Principle: Static surface-wave sensing
  • Diagnostics: Self-monitoring fault detection

Manufacturer reference ↗

Operation and integration

A surface wave travels within the stainless-steel probe to detect liquid without mechanical movement or vibration. The switch monitors itself, and its positive-safety principle makes a malfunction produce a detection state. Single- and dual-point arrangements can connect directly to an alarm system or CARLA. Small probe dimensions ease installation, while density, pressure and temperature changes do not require fluid-specific calibration.

Manufacturer reference ↗

Application and selection

Published uses include high-level tanker alarms, service tanks, bilge accumulation and bulk-carrier water ingress. The source emphasizes low mechanical wear and maintenance, but the sensor remains part of an engineered protection system. Alarm architecture and the chosen single/dual detection arrangement should match the vessel's protective function.

Manufacturer reference ↗

Case studies & further resources

Manufacturer Document

L-20 Lidec Level Alarm Switch for Limit Detection - Data Sheet - English

View resource ↗
Manufacturer Document

L-92 Lidec Level Alarm Switch for Limit Detection - Data Sheet - English

View resource ↗
Manufacturer Document

L92-C Lidec Level Alarm Switch - Chemical Version for Limit Detection - Data Sheet - English

View resource ↗
Manufacturer Document

L91-C Lidec Level Alarm Switch - Chemical Version for Limit Detection - Data Sheet - English

View resource ↗
Manufacturer Document

L-91 Lidec Level Alarm Switch for Limit Detection - Data Sheet - English

View resource ↗
Manufacturer Document

L20-70 Lidec Level Alarm Switch for Limit Detection - Data Sheet - English

View resource ↗
Manufacturer Document

L-WIDS Lidec Level Alarm Switch for Limit Detection - Data Sheet - English

View resource ↗

Specifications

Processing & Performance Metrics

Accuracy
± 2 mm
Maximum Pressure
25 bar

Standards & Certifications

Intrinsic Safety Approvals
ATEX, IECEx., II 1 G D, Ex ia IIC or IIB T6 or T5 or T4 Ga, Ex ia IIIC T80 deg C or T95 deg C or, T130 deg C Da
Protection Class
IP66, IP67

Environmental Conditions

Ambient Temperature Range
-25 to + 70 deg C
Product Temperature
-40 to + 150 deg C

Technical Specifications

Process Connections
Flange ISO PN20 DN40

Electrical Specifications

Output Options
8mA/16mA Current, 16mA/8mA Current, NC Relay Contact, NO Relay Contact
Power Supply
18 to 28 VDC

Mechanical & Physical Properties

Probe Lengths
160 to 3000 mm/126 to 3000 mm

Applications & Compatibility

Applications
Applications requiring no calibration and online proof testing.; High‑level & overfill alarms; Marine liquid level alarm detection for chemicals & high‑temperature products
Mounting
Vertical - On Deck

Material Specifications

Materials
316L/904L Stainless Steel
Lidec Liquid Level Switch — manufacturer page values
ParameterPublished value / configuration
Accuracy± 2 mm
Maximum Pressure25 bar
Intrinsic Safety ApprovalsATEX, IECEx., II 1 G D, Ex ia IIC or IIB T6 or T5 or T4 Ga, Ex ia IIIC T80 deg C or T95 deg C or, T130 deg C Da
Protection ClassIP66, IP67
Ambient Temperature Range-25 to + 70 deg C
Product Temperature-40 to + 150 deg C
Process ConnectionsFlange ISO PN20 DN40
Output Options8mA/16mA Current, 16mA/8mA Current, NC Relay Contact, NO Relay Contact
Power Supply18 to 28 VDC
Probe Lengths160 to 3000 mm/126 to 3000 mm
ApplicationsApplications requiring no calibration and online proof testing.; High‑level & overfill alarms; Marine liquid level alarm detection for chemicals & high‑temperature products
MountingVertical - On Deck
Materials316L/904L Stainless Steel

Values retain manufacturer model qualifications. Review Honeywell specifications ↗

Resources

Filter documents by type

Resources linked from the manufacturer’s page. Confirm the document revision and applicable model before installation or ordering.

Honeywell resource listing ↗

Part Number

Published part numbers

10 parts

Part NumberDescriptionEnquiry
21243-HMSCOATED LIDEC 4W M10138 CONFIGURATEDRequest quotation
21419-HMSCOATED LIDEC M12027 2W CONFIGURATEDRequest quotation
29787A EXPLOSION PROOF SIREN DB3 (12-48V)V/A2Request quotation
33226KIT BACK INSTAL L20-70D CARGORequest quotation
34125-1ATEX. FALSH LIGHT RED XB12B02406RNBNNNNRequest quotation
34125-HMSATEX FALSHING LIGHT RED XB12 24VCC V A2Request quotation
34126-HMSATEX FALSHING LGHT ORANGE XB12 24VDC V ARequest quotation
34325-HMSCARLA PANEL 16 L92-230VRequest quotation
35125-10L20-70 LIDEC PROBE CABLE LENGTH 10mRequest quotation
35125-20L20-70 LIDEC PROBE CABLE LENGTH 20mRequest quotation
Manufacturer sources

https://process.honeywell.com/us/en/products/marine/fixed-gauging/lidec-liquid-level-switch

Last source check: 2026-11-02

Limit Switch Product Image New · Manufacturer ownership retained; republication licence not independently confirmed

Lidec Liquid Level Switch image · Manufacturer ownership retained; republication licence not independently confirmed

Lidec Liquid Level Switch image · Manufacturer ownership retained; republication licence not independently confirmed

Lidec Liquid Level Switch image · Manufacturer ownership retained; republication licence not independently confirmed

Lidec Liquid Level Switch image · Manufacturer ownership retained; republication licence not independently confirmed

Lidec Liquid Level Switch image · Site owner attests manufacturer permission; licence not independently audited

ABID Automation LTD maintains this independent product reference. Honeywell owns its product names and trademarks. Current manufacturer documentation governs specification, availability and certification.

Your next project. A clear starting point.

Tell us about your facility and what you need to achieve.

CallSend enquiry