Thermal Conductivity
7866 Thermal Conductivity Analyzer
7866 determines gas concentration from thermal conductivity, using a sensing unit, receiver and power supply.
7866 determines gas concentration from thermal conductivity, using a sensing unit, receiver and power supply. Single-range versions address binary or selected multicomponent mixtures; the triple-range arrangement supports hydrogen-cooled generator monitoring, including purge-gas measurements. Continuous sample flow through the sensing system supplies concentration information for power and other industrial gas-analysis applications.
Features & Benefits
- Versions
- Single range and triple range
- Supports continuous gas-concentration measurement
Applications
- Hydrogen-cooled generators
- Industrial binary-gas analysis
Detailed capabilities
What Is It?
7866 analyzes a binary gas mixture or a calibrated constituent in a multicomponent mixture using thermal conductivity. Single-range and triple-range versions serve different duties. Triple-range use includes hydrogen purity in cooled generators and the stated CO2/air/hydrogen power-plant measurements; the single-range form supports other selected gas-analysis tasks.
How Does It Work?
A continuously flowing sample enters the sensing unit. Its electrical output reaches a separate receiver over the interconnecting cable, and the receiver provides current output for remote observation or recording. Sensing unit, control unit and power supply form the system. Calibration for a multicomponent mixture is explicitly required rather than every unknown mixture being resolved automatically.
What Problems Does It Solve?
Gas concentration information supports safe, efficient operation where incorrect composition can impair productivity or create hazardous conditions. The source links sensitivity, stability and response to its solid-state design. This is process-analysis information; the page does not establish a standalone certified explosion-prevention system or universal identification of every gas component.
Processing & Performance Metrics
- Accuracy
- ±2% of span (output signal) at reference conditions
- Sample Flow Rate Influence
- <±0.5% of span (0.2 to 4 cfh/100 to 2000 cc/min)
- Maximum Drift
- Zero/Span: ±2% of span per week
Environmental Conditions
- Atmospheric Pressure Influence
- ±0.01% of span per inch H₂O (±0.05% per mm Hg)
- Ambient Temperature Influence
- Sensing Unit: <1% Full Scale Over Range, Control Unit: ±0.01% per deg C
- Analyzer Type
- Single Range or Triple Range (CO₂ in Air, H₂ in CO₂, H₂ in Air)
Electrical Specifications
- Ambient Requirements
- RH: 90% max, Temperature: -10 to +50 deg C, Storage Temperature: 70 deg C max
- Sample Requirements
- Temp: -10 to +50 deg C (14 to 122 deg F), Flow: 0.2 to 4.2 cfh (100 to 2000 cc/min), Pressure: 37 mm Hg min
- Power Requirements
- Control Unit: 90 to 264 VAC (18 VA max) or 24 VAC/DC (12 VA max)
- Output Signal
- 0 to 20 mA or 4 to 20 mA (max load 800 Ω)
Measurement & Sensor Related (Measurement Capabilities)
- Reproducibility
- ±1% of span (24 hour)
- Repeatability
- ±0.3% of span (short term)
- Linearity
- ±2% of span for most standard ranges
Technical Specifications
- Connection Type
- 1/4 in OD Tubing (Sample/Reference), 1/2 in Conduit (Power)
I/O Capabilities
- Response Time
- CO₂: <2s initial, 24s (63%), 45s (90%), 80s (99%) millisecond; H₂: <1s initial, 13s (63%), 23s (90%), 40s (99%) millisecond
Applications & Compatibility
- Applications
- Hydrogen purity in generators, hydrogen in ammonia/nitrogen, hydrogen in hydrocarbons
Standards & Certifications
- Approvals
- FM/CSA Class I, II, III Div 1 (Triple Range std, Single Range opt), CE, IP65 Panel Front
| Parameter | Published value / configuration |
|---|---|
| Accuracy | ±2% of span (output signal) at reference conditions |
| Sample Flow Rate Influence | <±0.5% of span (0.2 to 4 cfh/100 to 2000 cc/min) |
| Maximum Drift | Zero/Span: ±2% of span per week |
| Atmospheric Pressure Influence | ±0.01% of span per inch H₂O (±0.05% per mm Hg) |
| Ambient Temperature Influence | Sensing Unit: <1% Full Scale Over Range, Control Unit: ±0.01% per deg C |
| Analyzer Type | Single Range or Triple Range (CO₂ in Air, H₂ in CO₂, H₂ in Air) |
| Ambient Requirements | RH: 90% max, Temperature: -10 to +50 deg C, Storage Temperature: 70 deg C max |
| Sample Requirements | Temp: -10 to +50 deg C (14 to 122 deg F), Flow: 0.2 to 4.2 cfh (100 to 2000 cc/min), Pressure: 37 mm Hg min |
| Power Requirements | Control Unit: 90 to 264 VAC (18 VA max) or 24 VAC/DC (12 VA max) |
| Output Signal | 0 to 20 mA or 4 to 20 mA (max load 800 Ω) |
| Reproducibility | ±1% of span (24 hour) |
| Repeatability | ±0.3% of span (short term) |
| Linearity | ±2% of span for most standard ranges |
| Connection Type | 1/4 in OD Tubing (Sample/Reference), 1/2 in Conduit (Power) |
| Response Time | CO₂: <2s initial, 24s (63%), 45s (90%), 80s (99%) millisecond; H₂: <1s initial, 13s (63%), 23s (90%), 40s (99%) millisecond |
| Applications | Hydrogen purity in generators, hydrogen in ammonia/nitrogen, hydrogen in hydrocarbons |
| Approvals | FM/CSA Class I, II, III Div 1 (Triple Range std, Single Range opt), CE, IP65 Panel Front |
Values retain manufacturer model qualifications. Review Honeywell specifications ↗
Resources
Resources linked from the manufacturer’s page. Confirm the document revision and applicable model before installation or ordering.
Honeywell resource listing ↗Part Number
Published part numbers
10 parts
| Part Number | Description | Enquiry |
|---|---|---|
| 056173 | PCB HTR/RELAY BD | Request quotation |
| 07866D02 | DIGITAL THRML COND. GAS ANALYZER | Request quotation |
| 07866DC2 | REPL. DIGITAL CONTROL UNIT ONLY | Request quotation |
| 07866DHC2 | REPL DIGITAL CNTRL UNIT HYDROGEN | Request quotation |
| 07866DHH2 | DIGITAL THERMAL COND. ANALYZER | Request quotation |
| 07866HHSS | REPLACEMENT SENSOR ASSY. | Request quotation |
| 07866SS2 | REPL. SENSOR ASSEMBLY 2 PORT | Request quotation |
| 07866SS4 | REPL. SENSOR ASSEMBLY 4 PORT | Request quotation |
| 144914 | PC BOARD | Request quotation |
| 31072925 | GAUGE HYDROGEN | Request quotation |
Manufacturer sources
Last source check: 2026-11-02
7866 Thermal Conductivity Analyzer IMAGE · Manufacturer ownership retained; republication licence not independently confirmed
ABID Automation LTD maintains this independent product reference. Honeywell owns its product names and trademarks. Current manufacturer documentation governs specification, availability and certification.
Your next project. A clear starting point.