Fire & Life Safety · Building Systems · Industrial Automation [email protected] +880 1347 561 256
Honeywell Process Measurement & Control

Dissolved Oxygen

DL5000 Equilibrium Probe

DL5000 measures dissolved oxygen using an equilibrium-based probe rather than relying on the oxygen diffusion rate alone.

DL5000 measures dissolved oxygen using an equilibrium-based probe rather than relying on the oxygen diffusion rate alone. PVC versions address wastewater, while stainless-steel arrangements serve high-purity power applications. Compatible DirectLine modules or UDA2182 provide the measuring system, and several mounting choices support installation without routine replacement of internal electrolyte or electrode components.

Models: DL5000

Features & Benefits

Probe forms
PVC or stainless steel
  • Reduces internal probe maintenance

Applications

  • Industrial and municipal wastewater
  • Power-generation high-purity water

Detailed capabilities

What Is It?

DL5000 measures dissolved oxygen with DL424/DL425 modules or UDA2182. PVC construction addresses industrial/municipal wastewater, while the stainless version with its fittings addresses high-purity power duties. Inline, flow-through and immersed arrangements plus cable-length choices adapt installation. The material/version choice remains important despite the shared equilibrium sensing principle.

Manufacturer reference ↗

How Does It Work?

The equilibrium method responds to oxygen partial pressure rather than relying on oxygen diffusion rate as the governing measurement. It replenishes oxygen consumed during sensing and uses permanent electrodes, electrolyte and membrane. The heavy membrane permits oxygen diffusion as gas while avoiding the periodic internal replacement work described for conventional probes. These claims concern the stated probe design, not an absence of external cleaning.

Manufacturer reference ↗

What Problems Does It Solve?

The source describes stable ppb/ppm performance with reduced sensitivity to flow change or fouling and minimal internal maintenance. Cleaning and application-appropriate mounting remain necessary. Wastewater and high-purity power measurements therefore use different physical configurations of the same low-maintenance sensing concept rather than one body suitable for every process.

Manufacturer reference ↗

Specifications

Electrical Specifications

Oxygen Consumption
Negligible

Processing & Performance Metrics

Maximum Flow Rate
300 mL/min (with Flow Chamber) gallons per minute

Material Specifications

Housing Material
PVC or 316 Stainless Steel

I/O Capabilities

Mounting Types
Immersion, Insertion, or Flow-Through
Response Time
90% in 60 sec (After Warm-Up) millisecond
Calibration
Air and sample

Dimensions & Weight

Dimensions
219 x 34 mm OD (8.62 x 1.315 in), 1 in NPT pipe size

Others

Operating Temperature Range
2 to 60 deg C; 35.6 to 140 deg F
Storage Temperature Range
2 to 60 deg C; 35.6 to 140 deg F
DL5000 Equilibrium Probe — manufacturer page values
ParameterPublished value / configuration
Oxygen ConsumptionNegligible
Maximum Flow Rate300 mL/min (with Flow Chamber) gallons per minute
Housing MaterialPVC or 316 Stainless Steel
Mounting TypesImmersion, Insertion, or Flow-Through
Response Time90% in 60 sec (After Warm-Up) millisecond
CalibrationAir and sample
Dimensions219 x 34 mm OD (8.62 x 1.315 in), 1 in NPT pipe size
Operating Temperature Range2 to 60 deg C; 35.6 to 140 deg F
Storage Temperature Range2 to 60 deg C; 35.6 to 140 deg F

Values retain manufacturer model qualifications. Review Honeywell specifications ↗

Resources

Filter documents by type

Resources linked from the manufacturer’s page. Confirm the document revision and applicable model before installation or ordering.

Honeywell resource listing ↗

Part Number

Published model / variant identifiers. A full ordering code requires the manufacturer configuration.

Published part numbers

1 parts

Part NumberDescriptionEnquiry
DL5PPMDL5000 DISSOLVED OXYGEN PROBERequest quotation
Manufacturer sources

https://process.honeywell.com/us/en/products/process-instruments/analytical-instruments-and-sensors/dissolved-oxygen/dl5000-equilibrium-probe

Last source check: 2026-11-02

DL5000 Equilibrium Probe Image · Manufacturer ownership retained; republication licence not independently confirmed

ABID Automation LTD maintains this independent product reference. Honeywell owns its product names and trademarks. Current manufacturer documentation governs specification, availability and certification.